Current Position:Home>>Kits>>ELISA Kits>>QuicKey ELISA>> QuicKey Human MAU(Microalbuminuria) ELISA Kit
Cat.No.:E-TSEL-H0005
Welcome to order from local distributors.
Add to cart Bulk requestFor research use only. Order now, ship in 3 days
Assay type | Sandwich-ELISA |
Format | 96T/48T/24 |
Assay time | 2.5h |
Detection range | 1.56-100 ng/mL |
Sensitivity | 0.30 ng/mL |
Sample type &Sample volume | urine; 50μL |
Specificity | This kit recognizes Human MAU in samples. No significant or interference between Human MAU and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are < 10%. |
Application | This ELISA kit applies to the in vitro quantitative determination of Human MAU concentrations in urine.Please consult technical support for the applicability if other biological fluids need to be tested. |
Item | Specifications | Storage conditions after test |
---|---|---|
Micro ELISA Plate (Dismountable) | 96T: 8 wells ×12 strips48T: 8 wells ×6 strips | 2-8℃, 1 month |
Reference Standard | 96T: 2 vials48T: 1 vial | Discard unused reconstituted standard and dilutions |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8℃ |
Biotinylated Detection Ab Working Solution | 1 vial, 6 mL | |
HRP Conjugate Diluent | 1 vial, 14 mL | |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | |
Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 48T: 1 vial, 60 μL |
2-8℃ (Protect from light) |
Substrate Reagent | 1 vial, 10 mL | |
Stop Solution | 1 vial, 10 mL | 2-8℃ |
Plate Sealer | 5 pieces | |
Product Description | 1 copy | |
Certificate of Analysis | 1 copy |
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Human MAU were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Human MAU were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (ng/mL) | 4.14 | 12.39 | 49.50 | 4.74 | 11.35 | 40.46 |
Standard deviation | 0.22 | 0.56 | 2.66 | 0.24 | 0.56 | 2.04 |
CV (%) | 5.31 | 4.52 | 5.37 | 5.06 | 4.93 | 5.04 |
The recovery of Human MAU spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
---|---|---|
Urine(n=8) | 97-104 | 101 |
Samples were spiked with high concentrations of Human MAU and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Database Links | SwissProt: P02768 Entrez Gene: 213 |
Research Area | Cancer, Cardiovascular, Metabolism, Developmental biology, Signal transduction, Stem cells |
1. Add 50μL standard or sample to the wells, immediately add 50μL Biotinylated Detection Ab working solution to each well. Incubate for 90 min at 37°C |
|
2. Aspirate and wash the plate for 3 times |
|
3. Add 100μL HRP conjugate working solution. Incubate for 30 min at 37°C. Aspirate and wash the plate for 5 times |
|
4. Add 90μL Substrate Reagent. Incubate for 15 min at 37°C |
|
5. Add 50μL Stop Solution |
|
6. Read the plate at 450nm immediately. Calculation of the results |